Radiotherapy

Why radiotherapy is effective.

Cancer cells are mutated cells with a reduced ability to repair DNA. Therefore it is harder for them to repair the damage caused by irradiation than for healthy cells.

Cancer cells divide more often, and it is precisely during cell division that cells are most vulnerable to radiation.

Physical foundations

1 newton = 1 N = 1 kg m s-2

F = m a

1 joule = 1 J = 1 newton meter = 1 N m = 1 kg m2 s-2

$$ 1\ watt = 1\ W = {1\ joule\over 1\ second} = J/s = 1\ J s^{-1}$$

P = E / t = E t-1

1 hertz = 1 Hz = 1 oscillation per second = s-1

1 coulomb = 1 C = 1 ampere second = 1 A s

Q = I t

$$ potential\ difference\ = {work\ done\ in\ electrical\ circuit\ \over charge\ passing\ through\ circuit}$$

The definition of the potential difference is given in different ways. I suggest this one:

«If the potential difference between two points in space is 1 V, then to do 1 J of work it is enough to move a charge of 1 C between them»

«A field with a potential difference of 1 V between two points will spend 1 J of energy moving a charge of 1 C»

«A field with a potential difference of 2 V between two points will spend 2 J of energy moving a charge of 1 C»

If we draw an analogy with gravity, a change from 1 V to 2 V is as if g from 9.8 m s-2 grew to 19.6 m s-2. To lift the same mass in such a field you have to do twice as much work.

« The more work a charge of 1 C does moving between two points in space, the higher the potential difference between them. »

Let me explain the analogy with gravity. The potential energy is given by the formula E = mgh, that is the more mass moved to the height h the more energy. Or the greater the height to which a certain mass is moved, the higher the energy. The mass here is the analogue of the charge. Consider the formula

$$ Potential\ difference\ = {mgh \over m} = gl$$

Such a potential difference is measured in J / kg

As we have understood, the greater the value of the product gl between two points, the more work has to be done to move the mass between them.

gl in a gravitational field has no special name, so its dimension is simply joule divided by kilogram.

In an electric field the charge plays the role of the mass, so joule has to be divided by coulomb. The resulting quantity has a special dimension, the volt

$$ 1\ volt\ = 1\ V = {1\ joule \over 1\ coulomb}$$
$$ work\ done = Q\cdot V = I\cdot t\cdot V$$
$$ 1\ eV (a\ unit\ of\ energy) = {1.602 \times 10^{-19}C \times 1 volt} = 1.602 \times 10^{-19}J$$
$$ 1\ MeV = 10^6 eV = {10^6 \times 1.602 \times 10^{-19}C \times 1 volt} = 1.602 \times 10^{-13}J$$

Capacitance is a property of a conductor that characterises its ability to store electric charge. Capacitance is defined as the ratio of the charge of the conductor to its potential. Capacitance is denoted by C.

$$ Capacitance\ C = {Charge\ Q\ stored\ in\ the\ capacitor \over Potential\ difference\ V\ across\ the\ capacitor}$$

The same in English

$$ capacity\ C = {charge\ Q\ stored\ on\ conductor\ \over potential\ V\ to\ which\ conductor\ is\ raised}$$
$$ Q = {C \cdot V}$$

Let us look at capacitance from another side.

Hence we already know that

$$ work\ done = Q\cdot V = I\cdot t\cdot V$$

Or

$$ Q = {work\ done\ \over V}$$

If you have an idea to combine these formulas as follows

$$ {work\ done\ \over V} = Q = C \cdot V\ \color{red} ^*$$
$$ {work\ done\ \over V} = C \cdot V\ \color{red} ^*$$

And then get something like

$$ C = {work\ done\ \over V^2}\ \color{red} ^*$$
$$ work\ done = C \cdot V^2\ \color{red} ^*$$

* - that is wrong, it is given as a basis for thought

Then you have to remember that although these transformations give capacitance equal to the work done divided by the square of the voltage, that is not so.

The fact was not taken into account that as soon as the capacitor starts discharging the potential difference between the plates will fall and the amount of energy will also change. Therefore in the simplest case you have to take not V but the average voltage between 0 and V, that is V/2

$$ work\ done = C \cdot V \cdot {V \over 2} = C \cdot {V^2 \over 2}$$

At the same time the well-known formula for the energy stored in a capacitor looks as follows.

$$ W = {Q^2 \over 2 \cdot C} = {Q \cdot U \over 2} = {C\ \cdot U^2 \over 2}$$

Where W is the energy and U is the potential difference. The formula coincides with the one we obtained above.

It can be derived from the following equations

$$ dA = U \cdot dq = {1 \over C} \cdot q \cdot dq $$
$$ W = {1 \over C} \cdot {\int_{0}^{Q} \cdot q \cdot dq} = {Q^2 \over 2 \cdot C} $$

That is

$$ C = {W\ \over 2 \cdot U^2}$$

The unit of capacitance is the farad

$$ farad = 1\ F = {1\ coulomb \over volt} = 1\ C\ /\ V$$

$$a = 7.2 {km \over hr}.$$

When $a \ne 0$, there are two solutions to \(ax^2 + bx + c = 0\) and they are $$x = {-b \pm \sqrt{b^2-4ac} \over 2a}.$$

Water phantom

In external beam radiotherapy (External Beam Radiotherapy) transverse and longitudinal dose measurements are made with a radiation detector to characterise the radiation beams from medical linear accelerators.

As a rule, an ionisation chamber and a water phantom (Water Phantom) are used to create these radiation dose profiles. Water is used because it is equivalent to human tissue.

A water phantom is a special device used to check how and to what extent the human body will be irradiated when using a particular source of radiation or a medical diagnostic or therapeutic device that uses such a source.

Design It is a cube or a rectangular parallelepiped filled with water. Often it is just a large aquarium.

Examples

A small stationary water phantom for absolute dosimetry of vertical high-energy photon and electron radiation.

A small stationary water phantom for absolute dosimetry of horizontal high-energy photon and electron radiation. The design of the phantom allows up to three ionisation chambers to be irradiated simultaneously.

The universal 3D water phantom is intended for measuring dose distributions during acceptance testing, during periodic testing as part of a quality assurance programme, or when filling treatment planning systems with dosimetric data.

Principle of operation

The water column models the human body almost perfectly, and its easy permeability makes it possible to measure the level of irradiation at different depths, predicting the consequences of irradiating the human body with one or another medical device, without conducting experiments on people themselves.

Output data

The result of measurements on a water phantom is usually a data set needed for further calculations of beam parameters based on the available algorithms.

An example of such measurements may be the dose at a certain depth with a fixed field size.

The dose profile along the beam diagonal may also be of interest. Because the diagonal is larger than the side, and also taking into account the fact that the beam diverges, the profile is usually measured along half of the diagonal.

Suppose that the distance from the source to the surface of the water phantom is 100 cm. The field of linear accelerators usually has a maximum size of 40 by 40 cm. This means the diagonal at the surface is 40 * the square root of 2 = 56.6 cm.

If we go down 30 cm in depth, this value must be multiplied by 1.3 = 73.6 cm. A water phantom of that size would have a considerable weight and would be somewhat more expensive. Half of this value is only 37 cm, which does not exceed the standard field size.

Gross Tumor Volume

GTV is the visible or palpable volume of a malignant tumour and its metastases. It usually corresponds to the densest part of the malignant tumour.

Clinical Target Volume

CTV includes GTV and the areas around the GTV in which diseased cells are suspected to be present.

The CTV is determined by the physician and depends on the dose. To determine it, only oncological/ anatomical factors are taken into account.

The CTV is determined before treatment planning and does not depend on how the treatment will be carried out. Possible tissue movements, for example during breathing, are not taken into account.

The concepts of GTV and CTV can be used both in External Beam Therapy and in surgery and in brachytherapy (Brachytherapy)

Planning Target Volume

This term has an obsolete name, Target Volume.

PTV is a geometrical concept defined in such a way as to choose suitable sizes of beams and other elements, taking into account the combined effect of all possible geometrical variations in such a way that CTV it is guaranteed to receive the dose prescribed by the physician.

Which factors must be taken into account:

Glossary

bilaterally - on both sides
dorsal - of the back
invasion - intervention, encroachment, penetration
laterality - lateralisation, asymmetry
lattice - grid
nasopharynx - nasal pharynx
sternomastoid - relating to the sternum and mastoid
ventral - ventral, abdominal
vertebra - a bone of the spine
vertebral - of the spine